AZI P/N 700-0099-L Last update July 2014 USER MANUAL JEROME® J405 MERCURY VAPOR ANALYZER July 2014 Firmware Version 1.34 ARI
AZI Customer Service 800-528-7411 or 602-470-1414 Page 10 of 64 4. PRINCIPLE OF OPERATION The heart of the J405 is the well-proven gold film sensin
AZI Customer Service 800-528-7411 or 602-470-1414 Page 11 of 64 The instrument’s microprocessor automatically re-zeroes the digital meter at the star
AZI Customer Service 800-528-7411 or 602-470-1414 Page 12 of 64 5. INSTRUMENT OPERATION J405 Main Screen Display The J405 Main Screen is depicted
AZI Customer Service 800-528-7411 or 602-470-1414 Page 13 of 64 J405 Back Panel Connections The above-labeled connections are available from the bac
AZI Customer Service 800-528-7411 or 602-470-1414 Page 14 of 64 J405 User Interface Main Menu Structure NOTE: The main menu also contains a select
AZI Customer Service 800-528-7411 or 602-470-1414 Page 15 of 64 A USB keyboard (AZI P/N: 990-0230) can be used to navigate the menu system and for da
AZI Customer Service 800-528-7411 or 602-470-1414 Page 16 of 64 Interval Samples on the: Examples: 1 minute 1-minute marks 12:00, 12:01, 12:02, 12:
AZI Customer Service 800-528-7411 or 602-470-1414 Page 17 of 64 REGEN menu The sensor saturation meter on the main screen (to the right of s. sat) i
AZI Customer Service 800-528-7411 or 602-470-1414 Page 18 of 64 SET INTERVAL If auto-sampling is being used (as set in SAMPLE MODE above), the J405
AZI Customer Service 800-528-7411 or 602-470-1414 Page 19 of 64 SAVE TO USB DRIVE Before selecting SAVE TO USB DRIVE, connect the target USB memory
AZI Customer Service 800-528-7411 or 602-470-1414 Page 2 of 64 JEROME® J405 Mercury Vapor Analyzer Operation Manual PROPRIETARY RIGHTS NOTICE This
AZI Customer Service 800-528-7411 or 602-470-1414 Page 20 of 64 SYSTEM menu The SYSTEM menu is password protected to prevent unwanted or accidental
AZI Customer Service 800-528-7411 or 602-470-1414 Page 21 of 64 VIEW SENSOR STATS Use VIEW SENSOR STATS to display the current saturation level of t
AZI Customer Service 800-528-7411 or 602-470-1414 Page 22 of 64 Daily Operations Before each day's use of the Jerome® J405, perform the followi
AZI Customer Service 800-528-7411 or 602-470-1414 Page 23 of 64 Sensor Regeneration Sensor regeneration is needed to clean the J405 sensor of any ac
AZI Customer Service 800-528-7411 or 602-470-1414 Page 24 of 64 Sample Mode This mode, used for standard operation, produces optimum accuracy with t
AZI Customer Service 800-528-7411 or 602-470-1414 Page 25 of 64 Survey Mode The survey mode takes samples and displays the result every 2 seconds au
AZI Customer Service 800-528-7411 or 602-470-1414 Page 26 of 64 Operating on AC Power or Generator • For stationary use, the J405 may be operated o
AZI Customer Service 800-528-7411 or 602-470-1414 Page 27 of 64 Medium Frequency user: Once per month usage. For this type of user, the instrument sh
AZI Customer Service 800-528-7411 or 602-470-1414 Page 28 of 64 When operating using the external battery pack, the J405 will shut down after twenty
AZI Customer Service 800-528-7411 or 602-470-1414 Page 29 of 64 Retrieving Data On instruments that have the USB Communications option, sample data
AZI Customer Service 800-528-7411 or 602-470-1414 Page 3 of 64 Table of Contents 1. UNPACKING THE INSTRUMENT ...
AZI Customer Service 800-528-7411 or 602-470-1414 Page 30 of 64 Instrument I/O Interface The J405 I/O port is a DB-25F connector on the back of the
AZI Customer Service 800-528-7411 or 602-470-1414 Page 31 of 64 To utilize the 4-20mA output, connect the positive wire of your 4-20mA system to pin
AZI Customer Service 800-528-7411 or 602-470-1414 Page 32 of 64 6. MAINTENANCE Preventive Maintenance Schedule To keep the Jerome® J405 operating
AZI Customer Service 800-528-7411 or 602-470-1414 Page 33 of 64 Flow System The Jerome® J405's flow system is the crucial link between the sens
AZI Customer Service 800-528-7411 or 602-470-1414 Page 34 of 64 AG Filter Check The increased sensitivity of the Jerome® J405 requires the replaceme
AZI Customer Service 800-528-7411 or 602-470-1414 Page 35 of 64 Internal Filters Part Number 1. Scrubber Filter (1 elbow, 1 straight barb) Z2600 393
AZI Customer Service 800-528-7411 or 602-470-1414 Page 36 of 64 Replacing the Battery Pack • Press the I/O button to turn the power OFF. • Unplu
AZI Customer Service 800-528-7411 or 602-470-1414 Page 37 of 64 7. CALIBRATION Factory Calibration Service The Jerome® J405's gold film senso
AZI Customer Service 800-528-7411 or 602-470-1414 Page 38 of 64 8. J405 TROUBLESHOOTING Symptom Possible Cause Solution Power Problems Instrume
AZI Customer Service 800-528-7411 or 602-470-1414 Page 39 of 64 Symptom Possible Cause Solution Low response or erratic readings after a long pe
AZI Customer Service 800-528-7411 or 602-470-1414 Page 4 of 64 Functional Test Procedure ...
AZI Customer Service 800-528-7411 or 602-470-1414 Page 40 of 64 Symptom Possible Cause Solution No USB Drive is connected or USB Drive is full.
AZI Customer Service 800-528-7411 or 602-470-1414 Page 41 of 64 Potential Interferences Potential interferences to the Jerome® mercury vapor analyze
AZI Customer Service 800-528-7411 or 602-470-1414 Page 42 of 64 9. ACCESSORIES & MAINTENANCE PARTS PART # ITEM DESCRIPTION Y405-0901 J405 Acc
AZI Customer Service 800-528-7411 or 602-470-1414 Page 43 of 64 990-0217 Hard Side Carry Case Includes a molded case with die cut foam rubber inserts
AZI Customer Service 800-528-7411 or 602-470-1414 Page 44 of 64 200-0143 Battery Pack Assembly 200-0170 Car Accessory Cable 990-0225 10-to-1 Dil
AZI Customer Service 800-528-7411 or 602-470-1414 Page 45 of 64 Z2600 3940 Filter, Chlorine Filter 2600 3039 Filter, Fritware, .25 inch (pack of 20
AZI Customer Service 800-528-7411 or 602-470-1414 Page 46 of 64 A2600 0902 Stopper Assembly A2600 0903 Syringe Assembly 2600 0022 Syringe Needle
AZI Customer Service 800-528-7411 or 602-470-1414 Page 47 of 64 10. APPENDIX A – USB/HYPERTERMINAL SETUP Installing the J405 USB Driver Note: The
AZI Customer Service 800-528-7411 or 602-470-1414 Page 48 of 64 3) Windows® may display one or more messages indicating that the driver has not pass
AZI Customer Service 800-528-7411 or 602-470-1414 Page 49 of 64 5) Power on the J405 and connect it to the target computer using the included USB ca
AZI Customer Service 800-528-7411 or 602-470-1414 Page 5 of 64 1. UNPACKING THE INSTRUMENT This manual contains details that will optimize the resu
AZI Customer Service 800-528-7411 or 602-470-1414 Page 50 of 64 7) Select “Install the software automatically”, and click Next. 8) Windows® may
AZI Customer Service 800-528-7411 or 602-470-1414 Page 51 of 64 9) Windows® will copy a file and finish the wizard. Click Finish. 10) Windows
AZI Customer Service 800-528-7411 or 602-470-1414 Page 52 of 64 Using HyperTerminal to receive data from the J405 1) The J405 should be connected t
AZI Customer Service 800-528-7411 or 602-470-1414 Page 53 of 64 5) Next, a window labeled Connect To will be displayed. Find the last dropdown list
AZI Customer Service 800-528-7411 or 602-470-1414 Page 54 of 64 7) A blank HyperTerminal screen will be displayed. On the status bar in the lower l
AZI Customer Service 800-528-7411 or 602-470-1414 Page 55 of 64 9) HyperTerminal will prompt for a file to save the data to. It is recommended that
AZI Customer Service 800-528-7411 or 602-470-1414 Page 56 of 64 12) When the J405 stops sending data, Click HyperTerminal's Transfer menu again
AZI Customer Service 800-528-7411 or 602-470-1414 Page 57 of 64 11. APPENDIX B – J405 FUNCTIONAL TEST KIT If your application requires frequent ver
AZI Customer Service 800-528-7411 or 602-470-1414 Page 58 of 64 Mercury Transfer • CAREFULLY pour the mercury into the center of the functional te
AZI Customer Service 800-528-7411 or 602-470-1414 Page 59 of 64 Syringe Technique • Pull and hold the syringe plunger against the bar-stop. • Ver
AZI Customer Service 800-528-7411 or 602-470-1414 Page 6 of 64 • Perform a sensor regeneration by following these steps: Connect the AC power sup
AZI Customer Service 800-528-7411 or 602-470-1414 Page 60 of 64 • When “Purging” disappears from the display, release the plunger and allow gravity
AZI Customer Service 800-528-7411 or 602-470-1414 Page 61 of 64 • Repeat the instructions for mercury injection three more times. The readings ob
AZI Customer Service 800-528-7411 or 602-470-1414 Page 62 of 64 Functional Test Troubleshooting If you do not achieve good results with the functio
AZI Customer Service 800-528-7411 or 602-470-1414 Page 63 of 64 12. WARRANTY Arizona Instrument LLC (seller) warrants to buyer that Jerome® product
AZI Customer Service 800-528-7411 or 602-470-1414 Page 64 of 64 TRADEMARK AND COPYRIGHT PROTECTION Jerome®, Arizona Instrument®, AZI® and the styli
AZI Customer Service 800-528-7411 or 602-470-1414 Page 7 of 64 2. INTRODUCTION The Jerome® J405 Mercury Vapor Analyzer is an ambient air analyzer w
AZI Customer Service 800-528-7411 or 602-470-1414 Page 8 of 64 Applications • Regulatory detection compliance • Regulatory clean up compliance • A
AZI Customer Service 800-528-7411 or 602-470-1414 Page 9 of 64 3. JEROME® J405 TECHNICAL SPECIFICATIONS Display Range* Units: µg/m3 Units: mg/m3 0