35 Default Settings Appendix EEG Default Settings
380
EEG Default Settings
BIS Default Settings
EEG Settings Factory Adult Factory Pedi Factory Neo
TP, SEF On
M D F, P P F, D e l t a , T h e t a ,
Alpha, Beta
Off
SEF Threshold 90 %
Numeric Average 8 sec
Wave Scale 100uV (or +/- 50uV if Show Gridlines is configured to Yes)
Show Gridlines No
Low Filter 0.5 Hz
High Filter 30 Hz
Impedance Limit 5 kOhm
Smoothing CSA On
Color Yellow
BIS Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
SQI On
EMG On
SR On
SEF Off
TP Off
Bursts On Off
Scale 100uV (or +/- 50uV
if Show Gridlines
is configured to Yes)
Show Gridlines No
Filters On
Low Filter 2 Hz
High Filter 70 Hz
Notch Filter On
High Alarm Limit 70
Low Alarm Limit 20
Alarms On
Smoothing Rate 30 sec 15 sec
Color Yellow Magenta