EASI ECG Lead Placement 6 ECG, Arrhythmia, and ST Monitoring
103
EASI Electrode Placement
1 E (V)
on the lower sternum at the level of
the fifth intercostal space
2 A (LL)
on the left midaxillary line at the
same level as the E electrode
3 S (LA)
on the upper sternum
4 I (RA)
on the right midaxillary line at the
same level as the E electrode
5 N
reference electrode - can be anywhere,
usually below the sixth rib on the
right hip
1
2
3
4
5